Contact SupplierSpecifiaction | Maximum permissible systematic error | Maximum permissible random error | |||||
Code | Capacity (μl) | Increment (μl) | Test Volume (μl) | Inaccuracy (±) | Inaccuracy (±) | Imprecision (±) | Inaccuracy (±) |
E11145 | 0.1 - 2.5 | 0.05 | 2.5 1.25 0.25 |
2.50% 3.00% 12.00% |
0.062μl 0.037μl 0.03μl |
2.00% 3.00% 6.00% |
0.05μl 0.0375μl 0.015μl |
E11107 | 0.5-10 | 0.1 | 10 5 1 |
1.00% 1.50% 2.50% |
0.1μl 0.175μl 0.025μl |
0.80% 1.50% 1.50% |
0.08μl 0.075μl 0.015μl |
E11108 | 2-20 | 0.5 | 20 10 2 |
0.90% 1.20% 3.00% |
0.18μl 0.12μl 0.06 |
0.40% 1.00% 2.00% |
0.08μl 0.10μl 0.04μl |
E11146 | 5 - 50 | 0.5 | 50 25 5 |
0.60% 0.90% 2.00% |
0.3μl 0.225μl 0.1μl |
0.30% 0.60% 2.00% |
0.15μl 0.15μl 0.1μl |
E11109 | 10-100 | 1 | 100 50 10 |
0.80% 1.00% 3.00% |
0.80μl 0.50μl 0.30μl |
0.15% 0.40% 1.50% |
0.15μl 0.20μl 0.15μl |
E11110 | 20-200 | 1 | 200 100 20 |
0.60% 0.80% 3.00% |
1.20μl 0.80μl 0.60μl |
0.15% 0.30% 1.00% |
0.30μl 0.30μl 0.20μl |
E11147 | 50 - 200 | 1.1 | 200 100 50 |
0.60% 0.80% 1.00% |
1.20μl 0.80μl 0.50μl |
0.15% 0.30% 0.40% |
0.15% 0.30% 0.2% |
E11111 | 100-1000 | 5 | 1000 500 100 |
0.60% 0.70% 2.00% |
6μl 3.50μl 2μl |
0.20% 0.25% 0.70% |
2μl 1.25μl 0.70μl |
E11134 | 1000-5000 | 50 | 5000 2500 1000 |
0.50% 0.60% 0.70% |
25μl 15μl 7μl |
0.15% 0.30% 0.30% |
7.5μl 7.5μl 3μl |
E11135 | 2000-10000 | 100 | 10000 5000 2000 |
0.60% 1.20% 3.00% |
60μl 60μl 60μl |
0.20% 0.30% 0.60% |
20μl 15μl 12μl |
Code | Capacity (μl) | No. per Case |
E11145 | 0.1 - 2.5 | 1 |
E11107 | 0.5-10 | 1 |
E11108 | 2-20 | 1 |
E11146 | 5 -50 | 1 |
E11109 | 10-100 | 1 |
E11110 | 20-200 | 1 |
E11147 | 50 - 200 | 1 |
E11111 | 100-1000 | 1 |
Large Volume Single Channel Adjustable Premium Plus Pipettes | ||
Code | Capacity (μl) | No. per Case |
E11134 | 1 - 5 | 1 |
E11135 | 2 - 10 | 1 |