We offer a complete product range of BIOHAZARD SAFETY CABINETS and Biological Safety Cabinets

BIOHAZARD SAFETY CABINETS

Apart from performing the functions of the regular laminar airflow equipment, it also protects the operator from the toxic / pathogenic contamination created during the research activities and also protects the atmosphere from such contaminations. For example, if an operator performs research on HIV virus or any other very hazardous / toxic materials, the operator needs to protect from the possible contamination /infection due to the virus he handles in his laboratory. A bio-safe hood does the job of protecting the not only the operators but also the environment by removing those viruses from the air-stream by means of High Efficiency Particulate Air Filters.

The biological safety cabinets (BSCs), enclosed containers and other engineering controls designed to remove or minimize exposures to hazardous biological materials. The biological safety cabinet (BSC) is the principal device used to provide containment of infectious splashes or aerosols generated by many microbiological procedures. Three types of biological safety cabinets (Class I, Class II (A1, A2, B1, & B2 ), Class III used in microbiological laboratories application. Open-fronted Class I and Class IIbiological safety cabinets are primary barriers which offers significant levels of protection to laboratory personnel and to the environment when used with good microbiological techniques. The Class II biological safety cabinet also provides

Biological Risk Assessed   Protection Provided

BSC CLASS

Personnel Product Environment
BSL 1-3 YES NO YES CLASS I
BSL 1-3 YES YES YES II TYPE A1, A2, & B2
BSL 1-4 YES YES YES III B1, B2

 

Type

Face Velocity(1fpm)

Airflow Pattern Radio-nuclides / Toxic Chemicals Bio-safety Level(s) Product Protection
Class I Open Front 75 In at front; rear and top through HEPA filter NO 2,3 YES

Class IIType A

75 In at front; rear and top through HEPA filter NO 2,3 YES
Type B1 100 Supply air 30% re-circulated through HEPA; remaining 70% exhaust through HEPA and hard ducted to atmosphere Yes (Low levels / Volatility) 2,3 YES
Type B2 100 No re-circulation; total exhaust via HEPA and hard ducted YES 2,3 YES
Type A2 100 Same as type IIA, but plenum under negative pressure to room and exhaust air is ducted to atmosphere YES 2,3 YES

Class III B 1 &B2

NA Supply air inlets and exhaust through 2 HEPA filter YES 3,4 YES