Technical Specifications:
Power Supply: AC100240V, 50/60HZ, Adapter: Output 9VDC, 5A
Battery:2500mAh 7.4V Rechargeable battery. Rated Charging Current:1250mA. Charge Cycle : 300 times
Working time:2hours
ECG: Lead Mode:3/5 Leads, I, II, III orI, II, III, AVR, AVL, AVF, V.Gain:2.5, 5.0mm/mV, 10mm/mV, 20mm/mV
Heart Rate: Measure Range : Adult : 15 300 bpm; Neonatal/Pediatric : 15350bpm, Accuracy : 1% Resolution :1 bpm; Differential Input Impedance> 5 M ohm
CMRR: Diagnostic Mode>90 dB; Monitor Mode >105 dB; Surgery Mode >105 dB; Leakage Current< 10 A
PACE Pulse Detects:Range : 2700mV ; Width : 0.12ms ; Rise Time : 10100s
Bandwidth:Surgery:1 20 Hz; Monitor : 0.5 35 Hz; Diagnostic : 0.05 130 Hz
ST Measurement: Range : -2.0 +2.0 mV ; Accuracy : -0.8mV+0.8mV: 0.02mV or 10%, which is greater
Other Range:unspecified
Respiration:Method Impedance between RA-LL, RA-LA; Differential Input Impedance>2.5 M ohm;
Respiration Impedance Range:0.33; Base Impedance Range : 200-4000; Bandwidth 0.3 2.5 Hz
Gain *0.25, *0.50, *1, *2,
Respiration Rate :Adult : 0 120 BrPM; Neonatal /Pediatric:0 150 BrPM; Resolution:1 BrPM; Accuracy:2 BrPM or 2%, use the greater. Apnea Alarm:10 40 s
NIBP:Method Oscillometric; Measure Mode : Manual, Auto, STAT; Measure Interval in AUTO Mode 199 min; Measure and Alarm Range : 0mmHg0kPa300mmHg40.0kPa; Static Pressure Accuracy : 3mmHg
Resolution: 1mmHg;
Accuracy: Maximum Mean error : 5mmHg; Maximum Standard Deviation : 8mmHg
Overpressure Protection: Adult : 300 mmHg;Pediatric : 240mmHg;Neonatal : 150 mmHg
SpO2:Measurement Range : 0 100 % ; Resolution : 1 %;
Accuracy:
| 70% 100% | 2 % |
| 40% 69% | 3 % |
| 0% 39% | unspecified |
Pulse Rate: Measure and Alarm Range : 20300bpm; Resolution : 1bpm; Accuracy : 3bpm
Temperature:Measure and Alarm Range : 0 50 C; Resolution : 0.1C
CO2(optional):MeasureRange : 0% 13%; Resolution : 1 mmHg; Breath Rate : 0 - 150 bpm
Recorder(optional):Paper Width : 50mm; Speed : 12.5/25/50 mm/s; WaveChannel : 3channels
Dimension & Weight:210 130 75mm; About 1Kg
Technical Specifications:
Display :12.1 inch LED (0-90 degree rotation)
Resolution :1024*768 pixels
Power supply and battery :100 - 240 VAC, 50 / 60Hz; Input Power: 70 to 100 VA
Battery :Rechargeable lithium battery : 14.8 V / 2200 mAh; Charge Cycle : more than 500 times
Working time :2 hours (optional larger battery for 4-5 hours)
Recorder (Built-in) :Thermal printer; Paper : Z-fold, thermo sensitive; Paper width: 112mm or 150mm; Print speed : 1/2/3 cm/min ; Trace : Max. 4 tracks ; Recording way : Real-time Recording ; Review printing
Alarm :Level : Low, Medium and High ; Indication : Auditory, Visual, delay of FHR alarm
Input Device :Knob, Keypad Input. Optional : Touch screen; Mouse / keyboard input
Trend & Reviewing :Cardiotocography (CTG) : Waveform review : 8640h; Vital signs: Trend graph and table : 2160h ; Waveform Review: 2h; Alarm Review : 2000 groups
Dimension and weight :360 * 310 * 120mm; 5.3KG (excluding accessories)
NST Report (Option)
Based on Fischer evaluation method (monitoring time 20 min)
Automatic Score result :
| Score | 8 - 10 excellent |
| Score | 5 - 7 suspicious |
| Score | 0 - 4 foetus hypoxia |