Features
Ultra-stable detector eliminates drift and need for constant readjustment., Built-in sample preparation.
The FlowSorb III measures surface area using the flowing gas method, which involves the continuous flow of an adsorptive and inert gas mixture over the sample at atmospheric pressure. The FlowSorb III 23052310 provides single-point and multipoint BET surface area measurements with speed, accuracy and reliability. Samples having surface areas from 0.01 m2g to over 1, 000 m2g are easily accommodated. The FlowSorb IIIs reproducibility is better than 0.5%. Confirmation of performance is desirable from time to time for every analytical instrument. The need for validation can arise from a change of operators, a peculiar result, an unfamiliar material, or simply from equipment wear and tear. Micromeritics products, with a few exceptions, do not require calibration per se; they derive results from well-established laws utilizing detection of basic parameters such as time, temperature, pressure, mass, and the like. Micromeritics instruments, nevertheless, are no exception to the desirability for periodic performance evaluation, because detectors, like everything else, age, drift, and occasionally fail.