Min. Order (MOQ)
1 piece
-
Driven Type
Electric
-
Automatic Grade
Manual
Technical specificationspipeline inlet pressurecylinder inletcylinder pressure reducerflowmeter rangesafetyoxygen flushchangeover unitbreathing circuitdimensionmodelsoptional accessorieso2n2oair2.7-6 bar2.7-6 bar2.7-6 bar(optional)four pin index yokestwo each for oxygen and nitrous oxide cylindersindividual pressure reducer for each yoke followedby a common second stage for stable working pressure.O2n2oair0-10 ltrs.min.0-12 ltrs.min.0-12 ltrs.min. (optional)orc (optional)nitrolockoxyfailure alarmpressure relief valveoxygen ratio controller ensuresminimum 25% oxygen at all time.Model pedias c (orc).Basic safety against accidental deliveryof nitrous oxide.When oxygen owmeteris closed. Model-pedias c (nl)audible alarm for minimum sevenseconds in case of oxygen supplypressure drops below 2k pa x 100.Mechanical safety against increase in low pressure circuit set at 20 kpa.Upto 55 ltrs. Of oxygen ow, with spring loaded push switchtoggle switch for changing from open to close circuit and viceversamagills breathing circuit-complete with angle mount, breathing hose, reservoir. Bag, expiratory valve & face