Our Products
Our offered Product range includes Humidity Chamber For LR Cubes and Environmental Test Chamber.
For maintaining a consistent temperature i.e. 50 ±2°C and Relative Humidity of minimum 90-95% ±3% in the mortar cubes, these cubes are to be kept in the Humidity Chamber for 24 hours before curing.
Manufactured using quality tested mild steel sheets with three coats of paints; the Environmental Test Chamber has reliable performance. We are reliable Manufacturer, Exporter and Supplier of Environmental Test Chamber from New Delhi, India. The offered Environmental Test Chamber is double walled and the interior chamber is insulated with 150 mm thick glass wool and puff insulation all around for limiting the radiation within the chamber only.
Construction
Air CirculationThe homogeneous condition of temperature inside the chamber maintaining by a forced blower with continuous duty rated TEFC motor for circulation. The conditioning space in the main chamber is baffled and heater and cooling coil is located in the conditioning space. The refrigeration unit is fitted under the chamber so as to ensure minimum space utilization.
Heating SystemLow surface loading sealed tubular heaters conforming to IS-4159 provided to add heat in the chamber. The heaters are located in the conditioning space and there will not be any direct radiation of heat on the item under test.
Humidity SystemLow-pressure moisture generator will be provided to generate moisture and add humidity in the chamber. The generator will be of minimum water level controller and mechanical float type refilling arrangement. Soft water will be used for the moisture generator.
Refrigeration SystemTwo-stage air cooled cascade refrigeration system will be provided working on freon 22 and 13 as refrigerants with evaporative cooling attachment for working in high summer ambient condition. The system is fitted with HP/LP cut off switch, overload protector, backup fuse protection etc. Low RH condition is maintained by smooth tube evaporation coil dehumidification arrangement.
InstrumentationsDual loop microprocessor programmer with direct RH display is provided to carry out steady state as well as cyclic damp heat test. Solid state relay proportional control for heaters is fitted.
Control PanelControl Panel will be provided at the lower side of the Chamber, which will house Microprocessor Based Programmer with direct RH display with Printing Facility, On/Off Switches, Fuses, Contactors, Indicating Lamps etc.
WiringAll the channel type wiring will be as per IS code.
Utilities
Safety protection
Power Supply : 230/415 Volts ±5 %, Single/Three Phase AC power supply.
Specifications
|
Performance |
As per IS : 9000, JSS : 55555 & JSS : 50101. |
|---|---|
|
Work space |
As per Customer's Requirement |
|
Temperature range |
- 50°C to 150°C or As per Customer's Requirement |
|
Humidity |
20 to 97 % RH 3 %. |
|
Heating |
Average rate of 1oC / minute (empty chamber). |
|
Cooling |
Average rate of 0.25oC / minute (empty chamber). |
|
Gradient in space |
Less than 1.5 oC. |