Our Products
We offer a complete product range of Tds Meter and CONDUCTIVITY METER:
Ideal for all basic water testing applications. The best value meter in the world!
FeaturesSpecifications Range:0 - 9990 ppm (mg/L) Resolution:1 ppm (0-999 ppm); 10 ppm (1000-9990 ppm) Accuracy:+/- 3% Conversion Factor:NaCl (avg. 0.5) Power source:2 x 1.5V button cell batteries (included) Battery life:Approx. 1000 hours of usage Dimensions:15.5 x 3.1 x 2.3cm (6.1 x 1.25 x 1 inches) Meter weight:56.7g (2 oz) Weight with package:76.5g (2.7 oz)
The meter is used on sites exposed to rain, dust, and robust handling. Basically, it comes in two parts i.e. Meter & Sensor.
Sensor: It is the sensing element of meter, which senses the conductivity and sends the signal to the meter.
Meter:This is the unit that processes the information received from the sensor and displays the actual conductivity in the line.
|
TROUBLE |
PROBABLE CAUSE |
ACTION |
|
DURING COMMISSIONING |
||
|
Fluctuating |
Sensor not dipped in the line properly |
Loop the outlet tubing such that the tee always remains |
|
Wrong wiring |
Connect the wires as per drawing |
|
|
DURING NORMAL SERVICE |
||
|
Display |
Sensor not dipped in the line .properly |
Loop the outlet tubing such that the tee always remains flooded with water |
|
Deposition on the sensor |
Remove the sensor and clean it with Polish paper |
|
|
No Display |
High Voltage |
Check MOV/ FUSE, if it is burnt then replace it with new one |
|
Frequent |
MOV short |
Replace MOV* |