Company Information
Ask for more detail from the seller
Contact SupplierM4000 Optical Emission Analyzer for Alloy Analysis
M4000 optical emission analyzer for alloy analysis creatively combined with world leading argon circulation technology enables the analysis of Fe, Al, Cu and other bases materials.Emission analyzer is an ideal solution to fast and accurate quantitative analysis near the furnace, metal material quality control and scientific research.
Optical System
 •Paschen-Runge optical system
 •Mult-CCD detectors with high resolution
 •Wavelength range: 175-520nm
 •Optics sealed argon circulation system
Spark Stand
 •Open sample clamp
 •Easily operated sample clamp adapted to different sample geometrics
 •Optimized argon circuit design
 •Easy to clean and maintenance
Excitation Source
 •Programmable pulse digital source
 •Maximum discharge frequency: 1000 Hz
 •Maximum discharge current: 400 A
 •Spark ignition pulse:1-14kV
 •Spark excitation pulse:20-230V
 •Arc excitation pulse:20-60V
 •Excitation Hole:13mm
Â
Argon Requirement
 •Purity requirement:99.999%
 •Input pressure:0.5MPa
 •Max 3.5Lmin in sparking
 •Average standby flow: 1Lmin
Â
Electrical and Environmental Requirement
 •AC220V±20V,50Hz
 •Max 400VA in sparking
 •Average standby power 50VA
 •Room temperature: 10℃~30℃
 •Storage temperature:0℃~45℃
 •Relative humidity: 20%~80%