Company Information
Ask for more detail from the seller
Contact SupplierZirconia o2 sensors
probe series - short housing
features
* zirconium dioxide (zro2 ) sensing elements
* long life, non-depleting technology
* integral heating element
* high accuracy
* requires an external interface board to operate
benefits
* no reference gas required
* no need for temperature stabilisation
* compact enclosure
technical specifications
heater voltage
operating (standard response) 4v dc +- 0.1v dc (1.7a)
standby 1.65v dc (0.7a)
operating (fast response) 4.35v dc +- 0.1v dc (1.85a)
standby 2v dc (0.85a)
pump impedance at 700°c
permissible gas temperature -100'c to +250'c
gas flow rate 0 - 10 ms
repetitive permissible acceleration 5g
incidental permissible acceleration 30g
output values
oxygen pressure range 2mbar - 3bar max
accuracy 5mbar max
internal operational temperature 700'c
response time (10-90% step)
standard response sensor
fast response sensor
warm up time (prior to sensor operation) 60s
warm up time (from standby) 20s
output stabilisation time 180s.