•Measurement of O₂ in laboratory experiments
or of Oâ‚‚ gradients in soil/underground
•Monitoring gaseous O₂ in indoor environments
for climate control, in compost piles, and in
mine tailings
•Monitoring redox potential in soils
•Determination of respiration rates through
measurement of Oâ‚‚ consumption in sealed
chambers
OUTPUT OPTIONS
•Analog V DC
•SDI-12
•or hand-held meter
HEATED DETECTOR
The gas permeable membrane inlet can be heated to prevent water vapor from condensing and blocking the diffusion path. When sensors are used in soil or compost.
Features
UNIQUE DESIGN
Sensors have an internal thermistor fully sealed for temperature monitoring. Sensors are housed in a polypropylene body for long-term use underground. Two head options include a diffusion head for field use and a flow-through head for laboratory use.
SIMPLE CALIBRATION
Voltage output is linearly proportional to absolute amount of oxygen. Calibration is accomplished by measuring the voltage under ambient conditions (atmosphere is 20.95 % Oâ‚‚) and deriving a linear calibration factor. A zero offset can be obtained with Nâ‚‚ gas (recommended for measurements below 10 % Oâ‚‚).