Company Information
Ask for more detail from the seller
Contact SupplierDescription Triple walled unit, the sterilizer consists of a chamber with steam jacket and outer insulation cover � all are made of heavy gauge SS304316 grade stainless steel Steam storage � Subsequent sterilization can be in quick succession as steam generated is retained even when the load chamber is opened Standard features: Chamber condensate line with solenoid valve, steam release valve, pressure gauge, compound gauge, thermometer, sight glass, silicone gasket Drying system � Mechanical ejector system for drying the working chamber Safety devices: Pressure switch, safety valves, vacuum breaker, low water level cut-off Working Pressure: 1.2 kgfcm & 2.1kgfcm Working Temperature: 121°C 134°C Power Supply: 440 V, 3 Phase, 50 Hz AC Supply (Optional: 230 V, 1 Phase, 50 Hz AC Supply) Hydrostatic Test: Hydraulically tested up to 2.5 times the working pressure Features of Automatic Sterilizer: Manufactured in accordance with IS:3829 (Part 3) standards � for applicable models
Automatic SIEMENS PLC based control system
Sterilization pressure, temperature and drying time can be adjusted
Automatic sterilization by pre-programming the cycle
Digital display for pressure, temperature, power failure, low water, working time, end of sterilisation