Our Products
Our offered Product range includes SI-203 Digital Single Display Flame Photometer, SI-206 Digital Duel Display Flame Photometer, Microprocessor Flame Photometers, SI-209 Microprocessor Flame Photometer and Flame Photometer.
We are catering DIGITAL FLAME PHOTOMETER to our clients based globally. These are available at optimum prices with customized solutions as per clients’ specifications. Owing to the quality standards, these are highly demanded and are at par with international quality standards.
Salient Features
Accessories
We are catering DIGITAL FLAME PHOTOMETER to our clients based globally. These are available at optimum prices with customized solutions as per clients’ specifications. Owing to the quality standards, these are highly demanded and are at par with international quality standards.
Salient Features
Accessories
Salient Features:
Microprocessor Based with Printer Interface
Auto Gas Cutoff
Auto Flame Failure Detection
Four Measurements in Single Aspiration
Setup Storage Facility
Sample Data Storage Facility
Curve Calibration Programmability Using Maximum of 5 Standards
Multiple Calibration Curve Saving Facility
Data Processing Through Curve Fitting Techniques, Least Square, Quadratic
Online Help Key
Direct Results in ppm and meq
Useful for Medical & Soil Testing
The sample solution is aspirated through an auto-miser. Sample, Air and the fuel are mixed in the mixing chamber which is then sprayed as a very fine mist into the flame. The color of the flame is changed depending upon the concentration of elements present. Radiations from the flame pass through the sensing system and specific narrow band interference filter which permits only the characteristic radiation to pass to the photo-detector. The output of the photo-detector is then processed by the microcontroller and the final results are displayed on graphical LCD display.
In both models 5-Point calibration facility and non-linearity correction is provided through automatic curve fitting software. Facility has been provided for re-standardization with a single standard, thereby, eliminating the need of the multiple standards once a calibration has been performed.
Salient Features:
Specification
| Mode | General | Serum | Urine |
|---|---|---|---|
| Range | Na: 0 - 100 ppm Upto 250 meq/l, 1:100 dilK: 0 - 100 ppm Upto 250 meq/l, 1:100 dilCa: 0 - 100 ppm Upto 250 meq/l, 1:100 dilLi: 0 - 100 ppm Upto 250 meq/l, 1:100 dil | Na: 0 - 200 1:100 dilK: 0 - 10 1:100 dil-Li: 0 - 2 1:5 dil | Na: 0 - 250 1:100 dilK: 0 - 100 1:100 dilCa: 0 - 10 1:2 dil- |
| Sensitivity | General: Na: 0.5 ppm, K: 0.5 ppm, Li: 0.5 ppm, Ca: 15 ppm | ||
| Filters | Na & K supplied (Ca & Li are optional at extra cost) | ||
| Resolution | 0.1 ppm/meq | ||
| Reproducibility | < 1% coefficient of variation for 20 consecutive samples using 10 ppm Na set as maximum standard | ||
| Curve Fit Accuracy | < 2% error when 3 ppm Na/K and 5 ppm Li are set as maximum standards | ||
| Display | 128 x 64 Graphical LCD Display | ||
| Average Time | In-built in software | ||
| Flame System | LPG and dry oil free air | ||
| Detector | Silicon photodiode | ||
| Calibration | Upto 5-Point calibration with curve fitting software | ||
| Gas Control | Adjustable with knobs | ||
| Flame Failure | Auto detection | ||
| Gas Cutoff | Automatic | ||
| Ignition System | Auto ignition | ||
| Atomizer | Axial flow type | ||
| Printer | Provision for attachment of dot matrix printer or HP laserjet/ inkjet (non-windows) printers with centronics parallel interface | ||
| RS-232C | Available (PC software at extra cost) | ||
| Power | 230 V + 10% AC, 50 Hz | ||
| Dimensions | 285 x 255 x 210 mm (L x B x H) (Approx.) | ||
| Weight | 7.5 Kg (Approx.) | ||
COMPRESSOR UNIT |
|||
| Air Supply | By oil free mini compressor unit with pressure regulator | ||
| Combustion Gas | LPG controlled by precision regulator | ||
| Power | 230 V + 10% AC, 50 Hz | ||
| Dimensions | 290 x 255 x 210 mm (L x B x H) (Approx.) | ||
| Weight | 8 Kg (Approx.) | ||