Our Products
We offer a range of oxygen analyzers, designed to increase safety or improve efficiency in applications from industrial gas production, monitoring inert blanketing gases to combustion efficiency control. We use non-depleting measurement technologies thermo-paramagnetic and zirconium oxide to ensure long-term reliability and low cost of ownership.
paramagnetic oxygen analyzer - 3000ma
model 3000ma paramagnetic percent oxygen analyzer is a versatile, microprocessor based instrument specifically designed to detect oxygen in a variety of gases. The 3000ma is a precise, cost effective instrument capable of measuring percent oxygen in as low as the 0-1% range. Drawing on the solid performance of teledynes highly successful 3000 series control platform, the 3000ma provides the user with a wide variety of features and options to satisfy virtually any application.
features
proven, robust paramagnetic sensor
three user-selectable ranges plus cal range
signal and range id output: isolated 4-20 madc
programmable autoranging
two fully adjustable concentration alarm set pointswith programmable relay functions, form c contacts, 3a resistive
calibration contact span/zero, form a normally opencontact, 3a resistive
self diagnostics with form c failure alarm contacts
full duplex rs-232c communication link
five digit oxygen concentration led display
2x20 alphanumeric vacuum fluorescent display for setup and diagnostics
stainless steel sample passages and fittings
universal power supply: 100-240 vac / 50-60 hz
remote calibration digital inputs
Model LXT-280 Transmitter is a robust and versatile instrument for measurement and process control of dissolved oxygen. This instrument is Factory Mutual Approved as intrinsically safe and explosion proof for use in Class I, II, and III Divisions 1 Groups A through G hazardous areas. The Model LXT-280 transmitters provide solutions to a variety of liquid analytic applications involving hazardous areas.
Features
Magnetic switches located behind a glass window can be activated without having to open the instrument.Menu driven backlit display with contrast adjustmentAutomatic Calibration. This allows calibration points to be defined upon initialization of the transmitter.An easy back to factory calibration menu allows the instrument to return to a predetermined factory calibration. The default parameters are: zero electrode offset at 0 ppm or 0% oxygen and-400 mV at full scale.4-20mA current output transmitted on the same 24 VDC power wiring. The current output is fully expandable to any range within the transmitters full operating range. The current output can be configured to be reverse acting.Temperature Conversion. Temperature can be displayed in either Fahrenheit or Celsius. The units can be toggled from one to the other in the field with the conversion calculation performed automatically.
The Model 311PC is a portable, intrinsically safe percent oxygen analyzer which can be operated without an external power source and reliably calibrated without the use of cumbersome, questionable, so-called certified calibration gases.
Features
Micro-Fuel CellReliable CalibrationIntegral Power SupplyCompact PackagingAnalysis Ranges
The Model 2750 Turbine Generator Gas analyzer is a portable microcontroller based three-channel analyzer designed to measure the
concentration of a single gas component in a binary mixture of gases. A special thermal conductivity cell responds to the difference in thermal conductivity between the components of a gas mixture and produces a signal output which is linearized over the specific range of interest. Each channel is separately linearized for enhanced accuracy. The cell is maintained at a fixed temperature to minimize any thermally induced error.
Analysis Ranges
The Model 2750 has 3 separate user selectable analysis ranges controlled by a 3-position rotary switch.
Standard analysis ranges include:0-100% hydrogen in CO20-100% air in CO280-100% hydrogen in air