The 6500B series of Precision Impedance Analyzers provide precise and fast testing of components at frequencies up to 120 MHz. Basic measurement accuracy is 0.05% making the instruments the best in the class.

Display formats available include series or parallel equivalent circuit.

An Equivalent Circuit Anlysis function is available . This allows modelling and curve fitting to various equivalent circuit models. Four types of three component and a four component model are available. The instrument will calculate the nearest equivalent circuit parameters for the measurement traces and revise the results. Alternatively the component parameters may be entered and the instrument will plot the frequency characteristics.

For single frequency measurements a meter mode is available.

Option feature Polar and Complex Plots & Multi Measurement Mode

Variable drive and bias levels

AC drive levels up to 1V or 20mA can be selected to evaluate components in realistic operating environments.

The /D1 DC bias option provides 0 to +40V dc bias voltage and 0 to +100mA dc bias current while the /D2 option provides -40V to +40V dc bias voltage.

External control

The GPIB interface is used to control the instrument and read back measured values for applications such as quality control or for archiving purposes.

An Ethernet interface similarly allows the instrument to be controlled and to send out data allowing it to be integrated into many test environments.

Wide range of interfaces

Four front panel BNC connectors permit three or four terminal connections with the screens at ground potential.

The 1J1011 component fixture, supplied as standard, ensures optimum performance when measuring a wide range of components of leaded components and devices.

1J1012 (2 Terminal) and 1J1014 (4 Terminal) fixtures allow connection to surface mount devices.

Protection against charged capacitors

  • Key Features
    • Precise high frequency impedance measurements
    • Characterize components from 20 to 120 MHz
    • Wide range of models with maximum frequencies from 5MHz (6505B) to 120MHz (65120B)
    • 0.05% basic measurement accuracy
    • 0 to +40V DC bias voltage and 0 to 100mA DC bias Current (Option)
    • 40V DC bias voltage (Option)
    • Comprehensive measurement functions
    • Easy to use with large TFT touch screen
    • Clear graphic displays aids measurements
    • Traditional LCR Meter mode for single and repetitive measurements
    • Resonance Search
    • Equivalanct Circuit Analysis (Option)
    • Intuitive user interface
    • Fully programmable over GPIB and LAN
    • Polar and Complex Plots available (option)
    • Multi Measurement Mode available (option)
    • Keyboard and mouse optional control
    • Competitively priced

Technical Specification
Measurement Function Capacitance (C)
Inductance (L)
Resistance (R)
Reactance (X)
Conductance (G)
Susceptance (B)
Dissipation Factor (D)
Quality Factor (Q)
Impedance (Z)
Admittance (Y)
Phase Angle (Ø)
Frequency Range 6505B 20Hz to 5MHz
6510B 20Hz to 10MHz
6515B 20Hz to 15MHz
6520B 20Hz to 20MHz
6530B 20Hz to 30MHz
6550B 20Hz to 50MHz
65120B 20Hz to 120MHzFrequency step size 0.1 mHz
Accuracy of set frequency ±0.005″%
AC Drive Level 10 mV to 1V rms (varies with frequency)
200 µA to 20 mA rms (varies with frequency)
Measurement Speeds 4 speeds selectable
Up to 20 measurements per second for test frequencies ≧100 Hz
Accuracy Dissipation factor – ±0.0005 (1+D2)
Quality factor – ±0.05 %( Q+1/Q)
Capacitance / Inductance / Impedance – ±0.05%
Accuracy varies with frequency, drive level and measured impedance
DC Bias Current (Option) /D1 0 to +100mA dc bias current and 0 to +40V dc bias voltage
/D2 -40V to +40V dc bias voltage
Binning (Option) 10 bins with absolute and percentage limits
25 way d-type interface connector
/B1 (Non-isolated) Comon 0V Bin outputs 0 to 5V(normal) with >10mA current sink capability
/B2 (isolated) Comon 24V input outputs 0 to 24V with >10mA current source capability
Input Specification Power supply – Input voltage 90 V AC to 264 V AC (Autoranging)
Mains frequency 47 to 63 Hz
Display 8.4″ VGA (640 x 480) color TFT with touch panel
Measurement Connections Four front panel BNC connectors permit two, three or four terminal connections with the screens at ground potential.
1J1011 Component Fixture (supplied as standard) allows connection to leaded components and devices.
Printer HP-PCL compatible graphics printing (local and network)
Centronics / parallel printer port, Epson compatible text/ticket printing (local)
Remote Trigger Rear panel BNC with internal pull-up, operates on logic low or contact closure
GPIB Interface External instrument control 24 pin IEEE 488 connector
USB Interface Two Unversal Serial Bus (USB) interfaces
USB 1.1 compliant
VGA Interface 15-way D-type connector to drive an external monitor in addition to the internal screen
Network Interface 10/100-BASETX Ethernet controller
RJ45 connector
Keyboard Interface Standard USB or PS/2 keyboard port. Instrument front panel remains active with keyboard plugged in
Mouse Interface Standard USB or PS/2 mouse port
Touch screen remains enabled when the mouse is connected