My ExportersIndia

For Buyer

For Seller

For Help

Data Loggers

We offer a complete product range of Bowie Dick Test Data Logger

Bowie Dick Test Data Logger

Ensuring correct sterilisation of equipment in hospitals, veterinary and dental surgeries is critical to patient care. Ebro provide a validation kit that logs temperatures at all parts of an autoclave and this EBI 16 - SL 1520 starter kit is specifically designed for routine checks of steam sterilizers using the industry standard Bowie-Dick data logger that performs in accordance with the norms EN 285 / ISO 17665

Together with the evaluation software Winlog.med the new EBI 16 data logger forms an easy to use and very reliable electronic measurement system. With it a comprehensive routine control of steam sterilizers can be performed using an alternative Electronic Bowie Dick Test according to DIN EN ISO 11140-4. In addition to the review of the steam penetration, the relevant sterilization parameters are controlled.

  • Reliable: clear, reproducible measurement results
  • Accurate: high-resolution graphical cycle display
  • Secure: digital data recording and storage
  • Easy: to use and evaluate

Bowie Dick TestThe EBI 16 delivers clear results during daily checks of the air evacuation test and steam penetration test according to DIN EN 17665 and DIN EN 285.

Early warning systemThe EBI 16 provides early identification of possible failures in steam-sterilizers. Even small quantities of residual air that doesn't lead to a failed Bowie Dick Test yet, are detected.

Vacuum checkThe EBI 16 allows a reasonable vacuum check also for sterilizers without pressure display according to DIN EN 285.

Verification of sterilization parametersThe EBI 16 checks the sterilization parameters such as compensation time, hold time, sterilization temperature and sterilization time according to DIN EN 285

View Complete Details
Tell Us What are you looking for? Will call you back

Contact Us